Solid
Secondary Rhodamine WT Standards
for the 10-AU
Preparing
primary Rhodamine WT standards for fluorometry can be troublesome
as well as time consuming. Handling of concentrated dye solutions
should be minimized since they stain and are difficult to remove.
Since Rhodamine WT is temperature sensitive and photosensitive,
stability of the final standard is always a concern. Solid Secondary
standards circumvent some of these concerns, especially in field
applications where there is little time and few resources to prepare
new standards or examine the stability of existing standards.
When conducting dye studies, solid secondary standards provide
a simple, stable, and less expensive alternative to multiple calibrations
with several dilutions of a primary standard. Once the instrument
is calibrated with a primary Rhodamine WT standard, the solid
secondary standard can be used effectively for calibration, with
only occasional verification of results with a primary standard.
Thus, solid secondary standards make it simple to check for instrument
drift and to perform routine recalibrations. Turner Designs now
offers solid secondary standards, sealed in a holder, that fit
directly into the 13 mm adaptor of the 10-AU or Model 10 Fluorometer.
Each holder contains two ready-to-use fluorometric standard concentrations:
one high-level and one low-level concentration equivalent.
Features of the Solid Secondary
Standards
10-AU: P/N 10-AU-904
- Saves Time and Money.
Preparation of primary standards are capital and labor intensive.
Purchasing premade primary standards each time you calibrate
is extremely costly. Traditional secondary standards are often
used to decrease costs, but can require special handling and
preparation time. Our solid secondary standards are stable and
ready to use upon arrival with no preparation needed.
- No Special Storage Conditions.
The solid secondary standards can be stored in a drawer
or on the benchtop. They are not photosensitive or temperature
sensitive. Degradation is minimal after years of environmental
exposure.
- Guaranteed. The
solid secondary standards are guaranteed for two years after
arrival in your lab.
Calibrating with the Solid Secondary
Rhodamine WT Standards:
- Calibrate your fluorometer with
a known Rhodamine WT standard.
- Insert the 13 mm adaptor if
you have not already done so.
- Insert the solid secondary Rhodamine
WT standard and record the reading for low concentration. Rotate
your standard and record the reading for high concentration.
Record these numbers on the log sheet included with your standard.
- Future instrument calibrations
or checks may be performed by using the solid secondary standards
and values obtained in step 3.
Total Rhodamine WT Concentration
Specifications:
High Fluorometric Standard: Approximately
200 ppb
Low Fluorometric Standard: Approximately
20 ppb
|